Reproducibility and Reliability of Repeated Quantitative Fluorescence Angiography

15 Citationer (Scopus)

Abstract

INTRODUCTION: When using fluorescence angiography (FA) in perioperative perfusion assessment, repeated measures with re-injections of fluorescent dye (ICG) may be required. However, repeated injections may cause saturation of dye in the tissue, exceeding the limit of fluorescence intensity that the camera can detect. As the emission of fluorescence is dependent of the excitatory light intensity, reduction of this may solve the problem. The aim of the present study was to investigate the reproducibility and reliability of repeated quantitative FA during a reduction of excitatory light.

OriginalsprogEngelsk
TidsskriftSurgical Technology International
Vol/bind31
Sider (fra-til)35-39
ISSN1090-3941
StatusUdgivet - 2017

Fingeraftryk

Dyk ned i forskningsemnerne om 'Reproducibility and Reliability of Repeated Quantitative Fluorescence Angiography'. Sammen danner de et unikt fingeraftryk.

Citationsformater