Abstract
Physical inactivity is important to address, and an objective way of measuring inactivity is by accelerometry. The objective of this study was to determine the reliability and construct validity of the SENS motion system to record physical activity and inactivity in patients with knee osteoarthritis. Participants with an age > 40 years and an average weekly pain above 0 on a numeric rating scale (0 = no pain, 10 = worst pain) were included. Participants had a total of two study visits and at each visit participants completed a standardized activity. Data from 24 participants were analysed. A mean agreement of 99% (SD 3%) for sedentary behaviour and a mean agreement of 97% (SD 9%) for active behaviour were found. The agreement for "walking" was 28% (SD 18%). Mean agreement between recordings on the two visits was 96% (SD 8%) for sedentary behaviour and 99% (SD 1%) for active behaviour. The SENS motion activity measurement system can be regarded as a reliable and valid device for measuring sedentary behaviour in patients with knee OA, whereas detection of walking is not reliable and would require further work.
Originalsprog | Engelsk |
---|---|
Tidsskrift | Arthritis |
Vol/bind | 2018 |
Sider (fra-til) | 6596278 |
ISSN | 2090-1984 |
DOI | |
Status | Udgivet - 2018 |