Quality assessment in in vivo NMR spectroscopy: III. Clinical test objects: design, construction, and solutions.

M.O. Leach, D.J. Collins, S Keevil, Ian John Rowland, M.A. Smith, O Henriksen, W.M. Bovee, F Podo

43 Citationer (Scopus)

Abstract

Based on the requirements of test protocols developed to evaluate clinical MRS single slice and volume localisation sequences, two clinical test objects, STO1 and STO2 have been developed. The properties of a range of potential construction materials have been assessed, demonstrating that the water/Perspex interface produced minimum susceptibility effects. The design of the objects has been evaluated in trials on different magnetic resonance instruments, with size and loading being adjusted to allow use on currently available equipment. Appropriate test solutions for 31P and 1H measurements have been developed and their properties evaluated
OriginalsprogEngelsk
TidsskriftMagnetic Resonance Imaging
Vol/bind13
Udgave nummer1
Sider (fra-til)131-137
ISSN0730-725X
StatusUdgivet - 1995

Fingeraftryk

Dyk ned i forskningsemnerne om 'Quality assessment in in vivo NMR spectroscopy: III. Clinical test objects: design, construction, and solutions.'. Sammen danner de et unikt fingeraftryk.

Citationsformater