Forskning
Udskriv Udskriv
Switch language
Region Hovedstaden - en del af Københavns Universitetshospital
Udgivet

Planar quadrature coil design using shielded-loop resonators.

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Multi-site benchmarking of clinical 13C RF coils at 3T

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. In vivo single-shot 13C spectroscopic imaging of hyperpolarized metabolites by spatiotemporal encoding

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Dynamic nuclear polarization and optimal control spatial-selective (13)C MRI and MRS

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  • A Stensgaard
Vis graf over relationer
The shielded-loop resonator is known to have a low capacitive sample loss due to a perfect balancing. In this paper, it is demonstrated that shielded-loop technology also can be used to improve design of planar quadrature coils. Both a dual-loop circuit and especially a dual-mode circuit may benefit from use of shielded-loop resonators. Observations in measurements agree with theory for both a dual-loop coil and a dual-mode coil. The coils were designed for use as transmit/receive coil for 1H imaging and spectroscopy at 4.7 T in rat brain.
Bidragets oversatte titelPlanar quadrature coil design using shielded-loop resonators.
OriginalsprogEngelsk
TidsskriftJournal of Magnetic Resonance
Vol/bind125
Udgave nummer1
Sider (fra-til)84-91
Antal sider8
ISSN1090-7807
DOI
StatusUdgivet - 1997

ID: 32549512