Forskning
Udskriv Udskriv
Switch language
Region Hovedstaden - en del af Københavns Universitetshospital
Udgivet

Patch-based generation of a pseudo CT from conventional MRI sequences for MRI-only radiotherapy of the brain

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Deep learning for identification of critical regions associated with toxicities after liver stereotactic body radiation therapy

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Individualized estimates of overall survival in radiation therapy plan optimization - A concept study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Modeling tumor control probability for spatially inhomogeneous risk of failure based on clinical outcome data

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Technical note: DoseMapper - A validated GUI based exact numerical modelling method of shielding in PET/CT facilities

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Diffusion MRI outlined viable tumour volume beats GTV in intra-treatment stratification of outcome

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Tadalafil may improve cerebral perfusion in small-vessel occlusion stroke-a pilot study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Dual-energy material decomposition for cone-beam computed tomography in image-guided radiotherapy

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Home-based aerobic exercise in patients with lacunar stroke: Design of the HITPALS randomized controlled trial

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  5. The DAHANCA 32 study: Electrochemotherapy for recurrent mucosal head and neck cancer

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

PURPOSE: In radiotherapy (RT) based on magnetic resonance imaging (MRI) as the only modality, the information on electron density must be derived from the MRI scan by creating a so-called pseudo computed tomography (pCT). This is a nontrivial task, since the voxel-intensities in an MRI scan are not uniquely related to electron density. To solve the task, voxel-based or atlas-based models have typically been used. The voxel-based models require a specialized dual ultrashort echo time MRI sequence for bone visualization and the atlas-based models require deformable registrations of conventional MRI scans. In this study, we investigate the potential of a patch-based method for creating a pCT based on conventional T1-weighted MRI scans without using deformable registrations. We compare this method against two state-of-the-art methods within the voxel-based and atlas-based categories.

METHODS: The data consisted of CT and MRI scans of five cranial RT patients. To compare the performance of the different methods, a nested cross validation was done to find optimal model parameters for all the methods. Voxel-wise and geometric evaluations of the pCTs were done. Furthermore, a radiologic evaluation based on water equivalent path lengths was carried out, comparing the upper hemisphere of the head in the pCT and the real CT. Finally, the dosimetric accuracy was tested and compared for a photon treatment plan.

RESULTS: The pCTs produced with the patch-based method had the best voxel-wise, geometric, and radiologic agreement with the real CT, closely followed by the atlas-based method. In terms of the dosimetric accuracy, the patch-based method had average deviations of less than 0.5% in measures related to target coverage.

CONCLUSIONS: We showed that a patch-based method could generate an accurate pCT based on conventional T1-weighted MRI sequences and without deformable registrations. In our evaluations, the method performed better than existing voxel-based and atlas-based methods and showed a promising potential for RT of the brain based only on MRI.

OriginalsprogEngelsk
TidsskriftMedical Physics
Vol/bind42
Udgave nummer4
Sider (fra-til)1596-605
Antal sider10
ISSN0094-2405
DOI
StatusUdgivet - apr. 2015

ID: 45443013