Objective and Subjective Outcome of a New Transcutaneous Bone Conduction Hearing Device: Half-year Follow-up of the First 12 Nordic Implantations

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Stereologic Investigation of Mastoid Air Cell Geometry: Volume, Surface Area, and Anisotropy

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Cochlear MRI Signal Change Following Vestibular Schwannoma Resection Depends on Surgical Approach

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Intratympanic Steroid for Menière's Disease: A Systematic Review

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Nordic results of cochlear implantation in adults: speech perception and patient reported outcomes

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Cochlear implant should not be absolute contraindication for electroconvulsive therapy and transcranial magnetic stimulation

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Developing a national e-learning course in otorhinolaryngology: the Danish experience

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Tumor miRNA expression profile is related to vestibular schwannoma growth rate

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

OBJECTIVE: To examine the objective and subjective outcome of a new transcutaneous bone conduction hearing device.

STUDY DESIGN: Prospective, consecutive case series.

PATIENTS: Twelve patients were implanted. Eight patients had a conductive/mixed (con/mix) hearing loss. Four had single sided deafness.

MAIN OUTCOME MEASURES: At half-year follow-up, aided and unaided sound field hearing was evaluated by 1) warble tone thresholds, 2) pure-tone average (PTA4), 3) speech discrimination score (SDS) in quiet, and 4) speech reception threshold 50% at 70 dB SPL noise level (SRT50%). Subjective outcome was evaluated by three questionnaires: 1) International Outcome Inventory for Hearing Aids, 2) Speech, Spatial and Qualities of Hearing Scale 12, and 3) a questionnaire on frequency and duration of use.

RESULTS: No major complications occurred. The mean aided PTA4 was lowered by 23dB. SDS was increased by 40% at 50dB, by 34% at 65dB, and by 12% at 80 dB SPL. SRT50% in noise improved 5.2 dB. 58% of the patients used the device daily and 83% at least 5 days a week. 50% used the device ≥ 8 hours and 75% ≥ 4  hours a day. Mean International Outcome Inventory for Hearing Aids score was 3.7, corresponding to beneficial outcome. In Speech, Spatial and Qualities of Hearing Scale 12, "quality of hearing" scored especially high. The con/mix hearing loss group showed larger benefit especially in SDS, SRT50% in noise and the subjective evaluations, whereas frequency and duration of use were similar.

CONCLUSION: This study on the first 12 Nordic patients implanted with a new transcutaneous bone conduction hearing device demonstrates significant objective, as well as subjective hearing benefit. Patient satisfaction was high, as was the frequency of use.

OriginalsprogEngelsk
TidsskriftOtology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology
Vol/bind37
Udgave nummer3
Sider (fra-til)267-75
Antal sider9
ISSN1531-7129
DOI
StatusUdgivet - mar. 2016

ID: 49660160