Not Removing the Glossy White Cover from Adhesive INVOS Neonatal Sensors Affects the Oxygenation Measurement

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Harvard

APA

CBE

MLA

Vancouver

Author

Bibtex

@article{5b6b114154a947c09907461c3e07751c,
title = "Not Removing the Glossy White Cover from Adhesive INVOS Neonatal Sensors Affects the Oxygenation Measurement",
abstract = "The randomized clinical trial, SafeBoosC III, evaluates the effect of treatment guided by cerebral tissue oximetry monitoring in extremely preterm infants. Treatment should be considered, when cerebral oxygen saturation (StO2) drops below a predefined hypoxic threshold. This threshold value differs between different brands of instruments. To achieve high external validity, in this pragmatic trial all commercially available cerebral tissue oximeters have been accepted, provided their specific hypoxic threshold value has been determined in phantom studies. Since most companies produce sensors with an adhesive surface on the patient-contacting side, in the phantom studies these sensors were applied according to the specifications, i.e., the glossy cover was removed from the sensor. However, since the skin of preterm infants is particularly fragile, some neonatologists keep this cover on the adhesive sensors, to avoid the risk of skin injury when removing the sensor. Therefore, the aim of this study was to determine whether keeping this cover on leads to different StO2 values. To evaluate the effect of the cover, we performed multiple deoxygenations in a blood-lipid phantom and compared an INVOS neonatal sensor (Medtronic), with and without the cover, to a reference oximeter (OxiplexTS, ISS). As expected, the relationship of the StO2 between the INVOS neonatal sensor and OxiplexTS was linear (r2 = 0.999) with and without cover, but the cover influenced the linear equation: StO2_INVOS_cover = 1.133*StO2_ISS + 7.1 as opposed to StO2_INVOS_nocover = 1.103*StO2_ISS + 12.0. Furthermore, the hypoxic SafeBoosC III threshold differed as well: 60.3% with cover and 63.8% without cover. In conclusion, keeping the adhesive cover on an INVOS neonatal sensor results in lower measured values. At the hypoxic threshold, this is more than 3% (from 60.3% to 63.8%), and therefore, if clinicians keep the cover on the sensor, they need to be aware of this difference.",
keywords = "Adhesives, Brain, Humans, Infant, Infant, Newborn, Infant, Premature, Oximetry, Oxygen, Spectroscopy, Near-Infrared, Neonates, Liquid phantom, Cerebral tissue oxygenation, Near-infrared spectroscopy, Safeboosc, Blood-lipid phantom, Brain oxygenation",
author = "Hansen, {Mathias L{\"u}hr} and Daniel Ostojic and Stefan Kleiser and Gorm Greisen and Martin Wolf",
year = "2021",
doi = "10.1007/978-3-030-48238-1_56",
language = "English",
volume = "1269",
pages = "353--357",
journal = "Advances in Experimental Medicine and Biology",
issn = "0065-2598",
publisher = "Springer New York LLC",

}

RIS

TY - JOUR

T1 - Not Removing the Glossy White Cover from Adhesive INVOS Neonatal Sensors Affects the Oxygenation Measurement

AU - Hansen, Mathias Lühr

AU - Ostojic, Daniel

AU - Kleiser, Stefan

AU - Greisen, Gorm

AU - Wolf, Martin

PY - 2021

Y1 - 2021

N2 - The randomized clinical trial, SafeBoosC III, evaluates the effect of treatment guided by cerebral tissue oximetry monitoring in extremely preterm infants. Treatment should be considered, when cerebral oxygen saturation (StO2) drops below a predefined hypoxic threshold. This threshold value differs between different brands of instruments. To achieve high external validity, in this pragmatic trial all commercially available cerebral tissue oximeters have been accepted, provided their specific hypoxic threshold value has been determined in phantom studies. Since most companies produce sensors with an adhesive surface on the patient-contacting side, in the phantom studies these sensors were applied according to the specifications, i.e., the glossy cover was removed from the sensor. However, since the skin of preterm infants is particularly fragile, some neonatologists keep this cover on the adhesive sensors, to avoid the risk of skin injury when removing the sensor. Therefore, the aim of this study was to determine whether keeping this cover on leads to different StO2 values. To evaluate the effect of the cover, we performed multiple deoxygenations in a blood-lipid phantom and compared an INVOS neonatal sensor (Medtronic), with and without the cover, to a reference oximeter (OxiplexTS, ISS). As expected, the relationship of the StO2 between the INVOS neonatal sensor and OxiplexTS was linear (r2 = 0.999) with and without cover, but the cover influenced the linear equation: StO2_INVOS_cover = 1.133*StO2_ISS + 7.1 as opposed to StO2_INVOS_nocover = 1.103*StO2_ISS + 12.0. Furthermore, the hypoxic SafeBoosC III threshold differed as well: 60.3% with cover and 63.8% without cover. In conclusion, keeping the adhesive cover on an INVOS neonatal sensor results in lower measured values. At the hypoxic threshold, this is more than 3% (from 60.3% to 63.8%), and therefore, if clinicians keep the cover on the sensor, they need to be aware of this difference.

AB - The randomized clinical trial, SafeBoosC III, evaluates the effect of treatment guided by cerebral tissue oximetry monitoring in extremely preterm infants. Treatment should be considered, when cerebral oxygen saturation (StO2) drops below a predefined hypoxic threshold. This threshold value differs between different brands of instruments. To achieve high external validity, in this pragmatic trial all commercially available cerebral tissue oximeters have been accepted, provided their specific hypoxic threshold value has been determined in phantom studies. Since most companies produce sensors with an adhesive surface on the patient-contacting side, in the phantom studies these sensors were applied according to the specifications, i.e., the glossy cover was removed from the sensor. However, since the skin of preterm infants is particularly fragile, some neonatologists keep this cover on the adhesive sensors, to avoid the risk of skin injury when removing the sensor. Therefore, the aim of this study was to determine whether keeping this cover on leads to different StO2 values. To evaluate the effect of the cover, we performed multiple deoxygenations in a blood-lipid phantom and compared an INVOS neonatal sensor (Medtronic), with and without the cover, to a reference oximeter (OxiplexTS, ISS). As expected, the relationship of the StO2 between the INVOS neonatal sensor and OxiplexTS was linear (r2 = 0.999) with and without cover, but the cover influenced the linear equation: StO2_INVOS_cover = 1.133*StO2_ISS + 7.1 as opposed to StO2_INVOS_nocover = 1.103*StO2_ISS + 12.0. Furthermore, the hypoxic SafeBoosC III threshold differed as well: 60.3% with cover and 63.8% without cover. In conclusion, keeping the adhesive cover on an INVOS neonatal sensor results in lower measured values. At the hypoxic threshold, this is more than 3% (from 60.3% to 63.8%), and therefore, if clinicians keep the cover on the sensor, they need to be aware of this difference.

KW - Adhesives

KW - Brain

KW - Humans

KW - Infant

KW - Infant, Newborn

KW - Infant, Premature

KW - Oximetry

KW - Oxygen

KW - Spectroscopy, Near-Infrared

KW - Neonates

KW - Liquid phantom

KW - Cerebral tissue oxygenation

KW - Near-infrared spectroscopy

KW - Safeboosc

KW - Blood-lipid phantom

KW - Brain oxygenation

UR - http://www.scopus.com/inward/record.url?scp=85105590671&partnerID=8YFLogxK

U2 - 10.1007/978-3-030-48238-1_56

DO - 10.1007/978-3-030-48238-1_56

M3 - Journal article

C2 - 33966242

VL - 1269

SP - 353

EP - 357

JO - Advances in Experimental Medicine and Biology

JF - Advances in Experimental Medicine and Biology

SN - 0065-2598

ER -

ID: 67597107