Abstract
Ten surgical patients were monitored with nasal and finger pulse oximetry (Nellcor N-200) for five study periods with alternating mouth and nasal breathing and switching of cables and sensors. Nasal pulse oximetry was found to overestimate arterial oxygen saturation by 4.7 (SD 1.4%) (bias and precision).
Bidragets oversatte titel | Nasal pulse oximetry overestimates oxygen saturation. |
---|---|
Originalsprog | Engelsk |
Tidsskrift | Anaesthesia (Oxford) |
Vol/bind | 45 |
Udgave nummer | 12 |
Sider (fra-til) | 1070-1071 |
Antal sider | 2 |
ISSN | 0003-2409 |
Status | Udgivet - 1990 |