Abstract
PURPOSE: The R2* relaxation rate constant has been suggested as a sensitive measure for iron accumulation. The aim of this multi-center study was to assess the inter-scanner and inter-subject variability of R2* mapping and to investigate the relationship between brain volume and R2* in specific structures. METHODS: R2* mapping was performed in 81 healthy subjects in seven centers using different 3 T systems. R2* was calculated from a dual-echo gradient echo sequence and was assessed in several deep gray matter structures. The inter-scanner and inter-subject variability of R2* was calculated by means of the coefficient of variation before and after correcting for age. RESULTS: Significant center effects were seen in some regions which get lost after age correction. The coefficient of variation for the inter-center variability was much lower (
Originalsprog | Engelsk |
---|---|
Tidsskrift | Magnetic Resonance in Medicine |
Vol/bind | 71 |
Udgave nummer | 3 |
Sider (fra-til) | 1103–1107 |
ISSN | 0740-3194 |
DOI | |