Abstract
OBJECTIVE: To document the long-term surgical and audiologic results of the implantation of a first-generation semi-implantable hearing aid for mixed hearing loss, consisting of a magnetic partial or total ossicular replacement prosthesis and an electromagnetic driver fitted in the ear canal. The short-term results have previously been published as excellent in both surgical and audiologic terms. To analyze and discuss reasons of treatment failure, as well as future perspectives.
STUDY DESIGN: Patient file review, follow-up otomicroscopic examination and audiometry, as well as patient interview. SETTING Tertiary referral center at a University hospital.
PATIENTS: Of nine patients operated on, six with a mixed hearing loss after chronic otitis media could be evaluated primarily. Two underwent implantation with a partial and four with a total ossicular replacement prosthesis.
MAIN OUTCOME MEASURES: Use of and satisfaction with the hearing aid, apparatus function and integrity, hearing without electromagnetic driver, prostheses containment, drum abnormalities, nonaudiologic ear symptoms, e.g., discharge.
RESULTS: All patients attended follow-up for a mean of 9.5 years after implantation (range 8.8-10 years). No patients used the semi-implantable hearing aid at follow-up. The mean semi-implantable hearing aid using time was 24 months (range 3-60 months). The main reasons for discontinuation of use were problems fitting the electromagnetic driver correctly in the ear canal and prosthesis extrusion or dislocation.
CONCLUSIONS: Despite excellent short-term surgical and audiologic results, it is concluded that these first long-term results of the first-generation electromagnetic semi-implantable hearing aids with ossicular replacement prosthesis for mixed hearing loss are disappointing. However, improvements in the surgical approach and in the reshaping, refinement, and overall improvement of the semi-implantable hearing aid could lead to satisfactory treatment of this group of patients.
| Originalsprog | Engelsk |
|---|---|
| Tidsskrift | Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology |
| Vol/bind | 23 |
| Udgave nummer | 6 |
| Sider (fra-til) | 904-11 |
| Antal sider | 8 |
| ISSN | 1531-7129 |
| DOI | |
| Status | Udgivet - nov. 2002 |
| Udgivet eksternt | Ja |
Fingeraftryk
Dyk ned i forskningsemnerne om 'Long-term results and experience with the first-generation semi-implantable electromagnetic hearing aid with ossicular replacement device for mixed hearing loss'. Sammen danner de et unikt fingeraftryk.Citationsformater
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS