Forskning
Udskriv Udskriv
Switch language
Region Hovedstaden - en del af Københavns Universitetshospital
Udgivet

Influence of positron emitters on standard γ-camera imaging

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Validation of a method for accurate and highly reproducible quantification of brain dopamine transporter SPECT studies

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Procedure guideline for brain perfusion SPECT using 99mTc radiopharmaceuticals 3.0

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. A Probabilistic Approach to Delineating Functional Brain Regions

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Granulomatøs dermatitis på PET

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Attempts to Target Staphylococcus aureus Induced Osteomyelitis Bone Lesions in a Juvenile Pig Model by Using Radiotracers

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. [99mTc]-labelled interleukin-8 as a diagnostic tool compared to [18F]FDG and CT in an experimental porcine osteomyelitis model

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

UNLABELLED: Combined PET and SPECT scanning can give supplementary information. However, activity from PET radionuclides can cause background counts and increased dead time in γ camera imaging (SPECT or planar) because the 511-keV photons can penetrate collimators designed for lower energies. This study investigated how to manage this issue, including what levels of PET radionuclides can be tolerated when a γ-camera investigation is performed.

METHODS: Different combinations of (68)Ga (PET radionuclide), (99m)Tc (low-energy radionuclide), and (111)In (medium-energy radionuclide) were scanned by a γ camera. Standard low-, medium-, and high-energy collimators were used with the γ camera. Dead time and counts near and distant from the sources were recorded.

RESULTS: Down scatter from 511 keV can give rise to a considerable number of counts within the (99m)Tc or (111)In energy windows, especially when the PET source is close to the camera head. Over the full camera head, the PET source can result in more counts per megabecquerel than the SPECT source ((99m)Tc or (111)In). Counts from the PET source were distributed over a large region of the camera head. With medium- and high-energy collimators, the sensitivity to the PET radionuclide was found to be about 10% of the sensitivity to (99m)Tc and about 20% of the sensitivity to (111)In, as measured within a 3-cm-radius region of interest.

CONCLUSION: If PET radionuclides of activity 1 MBq or higher are present in the patient at the time of SPECT, a medium-energy collimator should be used. Counts from PET sources will in SPECT usually be seen as a diffuse background rather than as foci. The thick septa of high-energy collimators may result in structure in the image, and a high-energy collimator is recommended only if PET activity is greater than 10 MBq.

OriginalsprogEngelsk