Forskning
Udskriv Udskriv
Switch language
Region Hovedstaden - en del af Københavns Universitetshospital
Udgivet

Inductive measurement and encoding of k-space trajectories in MR raw data

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Consensus-based technical recommendations for clinical translation of renal T1 and T2 mapping MRI

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Survey of 800+ data sets from human tissue and body fluid reveals xenomiRs are likely artifacts

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Image artifacts from MR-based attenuation correction in clinical, whole-body PET/MRI

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Data-driven separation of MRI signal components for tissue characterization

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Sensitivity and resolution improvement for in vivo magnetic resonance current-density imaging of the human brain

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. On the reconstruction of magnetic resonance current density images of the human brain: Pitfalls and perspectives

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Elevated body weight modulates subcortical volume change and associated clinical response following electroconvulsive therapy

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

OBJECTIVES: The objective of this study was to concurrently acquire an inductive k-space trajectory measure and corresponding imaging data by an MR scanner.

MATERIALS AND METHODS: 1D gradient measures were obtained by digital integration, regularized using measured gradient coil currents and recorded individually by the scanner concurrently with raw MR data. Gradient measures were frequency modulated into an RF signal receivable by the scanner, yielding a k-space trajectory measure from the cumulative phase of the acquired data. Generation of the gradient measure and frequency modulation was performed by previously developed custom, versatile circuitry.

RESULTS: For a normal echo planar imaging (EPI) sequence, the acquired k-space trajectory measure yielded slightly improved image quality compared to that obtained from using the scanner's estimated eddy current-compensated k-space trajectory. For a spiral trajectory, the regularized inductive k-space trajectory measure lead to a 76% decrease in the root-mean-square error of the reconstructed image.

DISCUSSION: While the proof-of-concept experiments show potential for further improvement, the feasibility of inductively measuring k-space trajectories and increasing the precision through regularization was demonstrated. The approach may offer an inexpensive method to acquire k-space trajectories concurrently with scanning.