Indirect Calorimetry in Mechanically Ventilated Patients: A Prospective, Randomized, Clinical Validation of 2 Devices Against a Gold Standard

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Harvard

APA

CBE

MLA

Vancouver

Author

Bibtex

@article{113700e9ce1348368912acaf6f3033b7,
title = "Indirect Calorimetry in Mechanically Ventilated Patients: A Prospective, Randomized, Clinical Validation of 2 Devices Against a Gold Standard",
abstract = "BACKGROUND AND AIMS: The 2 currently available indirect calorimeters, CCM Express Indirect Calorimeter (MedGraphics, St Paul, MN) and Quark RMR ICU Indirect Calorimeter (COSMED, Rome, Italy), have not been validated against a gold standard in mechanically ventilated patients. Our aim was to do so using a gold-standard, modified Tissot bell-spirometer method in mechanically ventilated patients who were hemodynamically, respiratory, and metabolically stable.METHODS: We studied 30 patients undergoing general anesthesia and major gynecological surgery. We measured oxygen consumption ([Formula: see text]O2) and resting energy expenditure (REE) in a randomized, sequential, crossover design with double determination of each device.RESULTS: Compared with the modified Tissot bell-spirometer, the CCM Express Indirect Calorimeter demonstrated a mean Δ-REE of +361 kcal/d, corresponding to a 31{\%} overestimation of energy requirements. Bland-Altman analysis for REE showed a mean (SD) bias of 384 (124) with limits of agreement 142-627 kcal/d. QUARK RMR ICU demonstrated a mean Δ-REE of 81 kcal/d, corresponding to a 7{\%} overestimation of energy requirements. Bland-Altman plot analysis showed a mean (SD) bias of 77 (167) with limits of agreement -249 to 404 kcal/d.CONCLUSIONS: The QUARK RMR ICU Indirect Calorimeter compared better with the gold standard for values of [Formula: see text]O2 and REE than did the CCM Express Indirect Calorimeter in mechanically ventilated patients who were circulatory and respiratory stable. Both indirect calorimeters had low precision.",
author = "Allingstrup, {Matilde Jo} and Jens Kondrup and Anders Perner and Christensen, {Poul Lunau} and Jensen, {Tom Hartvig} and Henneberg, {Steen Winther}",
note = "{\circledC} 2016 American Society for Parenteral and Enteral Nutrition.",
year = "2017",
doi = "10.1177/0148607116662000",
language = "English",
volume = "41",
pages = "1272--1277",
journal = "Journal of Parenteral and Enteral Nutrition",
issn = "0148-6071",
publisher = "Sage Science Press (US)",
number = "8",

}

RIS

TY - JOUR

T1 - Indirect Calorimetry in Mechanically Ventilated Patients

T2 - A Prospective, Randomized, Clinical Validation of 2 Devices Against a Gold Standard

AU - Allingstrup, Matilde Jo

AU - Kondrup, Jens

AU - Perner, Anders

AU - Christensen, Poul Lunau

AU - Jensen, Tom Hartvig

AU - Henneberg, Steen Winther

N1 - © 2016 American Society for Parenteral and Enteral Nutrition.

PY - 2017

Y1 - 2017

N2 - BACKGROUND AND AIMS: The 2 currently available indirect calorimeters, CCM Express Indirect Calorimeter (MedGraphics, St Paul, MN) and Quark RMR ICU Indirect Calorimeter (COSMED, Rome, Italy), have not been validated against a gold standard in mechanically ventilated patients. Our aim was to do so using a gold-standard, modified Tissot bell-spirometer method in mechanically ventilated patients who were hemodynamically, respiratory, and metabolically stable.METHODS: We studied 30 patients undergoing general anesthesia and major gynecological surgery. We measured oxygen consumption ([Formula: see text]O2) and resting energy expenditure (REE) in a randomized, sequential, crossover design with double determination of each device.RESULTS: Compared with the modified Tissot bell-spirometer, the CCM Express Indirect Calorimeter demonstrated a mean Δ-REE of +361 kcal/d, corresponding to a 31% overestimation of energy requirements. Bland-Altman analysis for REE showed a mean (SD) bias of 384 (124) with limits of agreement 142-627 kcal/d. QUARK RMR ICU demonstrated a mean Δ-REE of 81 kcal/d, corresponding to a 7% overestimation of energy requirements. Bland-Altman plot analysis showed a mean (SD) bias of 77 (167) with limits of agreement -249 to 404 kcal/d.CONCLUSIONS: The QUARK RMR ICU Indirect Calorimeter compared better with the gold standard for values of [Formula: see text]O2 and REE than did the CCM Express Indirect Calorimeter in mechanically ventilated patients who were circulatory and respiratory stable. Both indirect calorimeters had low precision.

AB - BACKGROUND AND AIMS: The 2 currently available indirect calorimeters, CCM Express Indirect Calorimeter (MedGraphics, St Paul, MN) and Quark RMR ICU Indirect Calorimeter (COSMED, Rome, Italy), have not been validated against a gold standard in mechanically ventilated patients. Our aim was to do so using a gold-standard, modified Tissot bell-spirometer method in mechanically ventilated patients who were hemodynamically, respiratory, and metabolically stable.METHODS: We studied 30 patients undergoing general anesthesia and major gynecological surgery. We measured oxygen consumption ([Formula: see text]O2) and resting energy expenditure (REE) in a randomized, sequential, crossover design with double determination of each device.RESULTS: Compared with the modified Tissot bell-spirometer, the CCM Express Indirect Calorimeter demonstrated a mean Δ-REE of +361 kcal/d, corresponding to a 31% overestimation of energy requirements. Bland-Altman analysis for REE showed a mean (SD) bias of 384 (124) with limits of agreement 142-627 kcal/d. QUARK RMR ICU demonstrated a mean Δ-REE of 81 kcal/d, corresponding to a 7% overestimation of energy requirements. Bland-Altman plot analysis showed a mean (SD) bias of 77 (167) with limits of agreement -249 to 404 kcal/d.CONCLUSIONS: The QUARK RMR ICU Indirect Calorimeter compared better with the gold standard for values of [Formula: see text]O2 and REE than did the CCM Express Indirect Calorimeter in mechanically ventilated patients who were circulatory and respiratory stable. Both indirect calorimeters had low precision.

U2 - 10.1177/0148607116662000

DO - 10.1177/0148607116662000

M3 - Journal article

VL - 41

SP - 1272

EP - 1277

JO - Journal of Parenteral and Enteral Nutrition

JF - Journal of Parenteral and Enteral Nutrition

SN - 0148-6071

IS - 8

ER -