Indirect Calorimetry in Mechanically Ventilated Patients: A Prospective, Randomized, Clinical Validation of 2 Devices Against a Gold Standard

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Gastric Residual to Predict Necrotizing Enterocolitis in Preterm Piglets As Models for Infants

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Enteral Autonomy and Days Off Parenteral Support With Teduglutide Treatment for Short Bowel Syndrome in the STEPS Trials

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Repeated Metabolic Balance Studies in Patients With Short Bowel Syndrome

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Effect of immunoglobulin G on cytokine response in necrotising soft-tissue infection: a post-hoc analysis

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Using machine learning for predicting intensive care unit resource use during the COVID-19 pandemic in Denmark

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. The effects of adding quinolones to beta-lactam antibiotics for sepsis

    Publikation: Bidrag til tidsskriftReviewForskningpeer review

  4. A new tool to assess Clinical Diversity In Meta-analyses (CDIM) of interventions

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

BACKGROUND AND AIMS: The 2 currently available indirect calorimeters, CCM Express Indirect Calorimeter (MedGraphics, St Paul, MN) and Quark RMR ICU Indirect Calorimeter (COSMED, Rome, Italy), have not been validated against a gold standard in mechanically ventilated patients. Our aim was to do so using a gold-standard, modified Tissot bell-spirometer method in mechanically ventilated patients who were hemodynamically, respiratory, and metabolically stable.

METHODS: We studied 30 patients undergoing general anesthesia and major gynecological surgery. We measured oxygen consumption ([Formula: see text]O2) and resting energy expenditure (REE) in a randomized, sequential, crossover design with double determination of each device.

RESULTS: Compared with the modified Tissot bell-spirometer, the CCM Express Indirect Calorimeter demonstrated a mean Δ-REE of +361 kcal/d, corresponding to a 31% overestimation of energy requirements. Bland-Altman analysis for REE showed a mean (SD) bias of 384 (124) with limits of agreement 142-627 kcal/d. QUARK RMR ICU demonstrated a mean Δ-REE of 81 kcal/d, corresponding to a 7% overestimation of energy requirements. Bland-Altman plot analysis showed a mean (SD) bias of 77 (167) with limits of agreement -249 to 404 kcal/d.

CONCLUSIONS: The QUARK RMR ICU Indirect Calorimeter compared better with the gold standard for values of [Formula: see text]O2 and REE than did the CCM Express Indirect Calorimeter in mechanically ventilated patients who were circulatory and respiratory stable. Both indirect calorimeters had low precision.

OriginalsprogEngelsk
TidsskriftJournal of Parenteral and Enteral Nutrition
Vol/bind41
Udgave nummer8
Sider (fra-til)1272-1277
ISSN0148-6071
DOI
StatusUdgivet - 2017

ID: 49283250