Abstract
We examine various independent component analysis (ICA) digital signal processing algorithms for estimating the arterial oxygen saturation (SpO2) as measured by a reflective pulse oximeter. The ICA algorithms examined are FastICA, Maximum Likelihood ICA (ICAML), Molgedey and Schuster ICA (ICAMS), and Mean Field ICA (ICAMF). The signal processing includes pre-processing bandpass filtering to eliminate noise, and post-processing by calculating the SpO2. The algorithms are compared to the commercial state-of-the-art algorithm Discrete Saturation Transform (DST) by Masimo Corporation. It is demonstrated that ICAMS and ICAMF perform up to 13% better than DST. PPG recordings are done with a reflective pulse oximetry sensor integrated in an Electronic Patch. This system is intended for patients with chronic heart and lung conditions.
Originalsprog | Engelsk |
---|---|
Tidsskrift | I E E E Engineering in Medicine and Biology Society. Conference Proceedings |
Vol/bind | 2009 |
Sider (fra-til) | 4039-44 |
Antal sider | 6 |
ISSN | 1557-170X |
DOI | |
Status | Udgivet - 1 jan. 2009 |