Improving quantitative dosimetry in (177)Lu-DOTATATE SPECT by energy window-based scatter corrections

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Calibrated image-derived input functions for the determination of the metabolic uptake rate of glucose with [18F]-FDG PET

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Determination of kidney function with 99mTc-DTPA renography using a dual-head camera

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. A comparison of different energy window subtraction methods to correct for scatter and downscatter in I-123 SPECT imaging

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. First experiences from Copenhagen with paediatric single photon emission computed tomography/computed tomography

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Human biodistribution and radiation dosimetry of the 5-HT2A receptor agonist Cimbi-36 labeled with carbon-11 in two positions

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Minimally invasive assessment of hepatic function in children with indocyanine green elimination: a validation study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Very Early Response Evaluation by PET/MR in Patients with Lung Cancer-Timing and Feasibility

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. SPECT/CT Atlas of Quality Control and Image Artefacts: IAEA Human Health Series No. 36, IAEA, Vienna

    Publikation: Bog/antologi/afhandling/rapportBogFormidling

Vis graf over relationer

PURPOSE: Patient-specific dosimetry of lutetium-177 ((177)Lu)-DOTATATE treatment in neuroendocrine tumours is important, because uptake differs across patients. Single photon emission computer tomography (SPECT)-based dosimetry requires a conversion factor between the obtained counts and the activity, which depends on the collimator type, the utilized energy windows and the applied scatter correction techniques. In this study, energy window subtraction-based scatter correction methods are compared experimentally and quantitatively.

MATERIALS AND METHODS: (177)Lu SPECT images of a phantom with known activity concentration ratio between the uniform background and filled hollow spheres were acquired for three different collimators: low-energy high resolution (LEHR), low-energy general purpose (LEGP) and medium-energy general purpose (MEGP). Counts were collected in several energy windows, and scatter correction was performed by applying different methods such as effective scatter source estimation (ESSE), triple-energy and dual-energy window, double-photopeak window and downscatter correction. The intensity ratio between the spheres and the background was measured and corrected for the partial volume effect and used to compare the performance of the methods.

RESULTS: Low-energy collimators combined with 208 keV energy windows give rise to artefacts. For the 113 keV energy window, large differences were observed in the ratios for the spheres. For MEGP collimators with the ESSE correction technique, the measured ratio was close to the real ratio, and the differences between spheres were small.

CONCLUSION: For quantitative (177)Lu imaging MEGP collimators are advised. Both energy peaks can be utilized when the ESSE correction technique is applied. The difference between the calculated and the real ratio is less than 10% for both energy windows.

OriginalsprogEngelsk
TidsskriftNuclear Medicine Communications
Vol/bind35
Udgave nummer5
Sider (fra-til)522-533
Antal sider12
ISSN0143-3636
DOI
StatusUdgivet - maj 2014

ID: 45001569