T1 - Impact of leaflet splitting on coronary access after redo-TAVI for degenerated supra-annular self-expanding platforms
AU - Beneduce, Alessandro
AU - Khokhar, Arif A
AU - Curio, Jonathan
AU - Giannini, Francesco
AU - Zlahoda-Huzior, Adriana
AU - Grant, Daire
AU - Lynch, Lisa
AU - Zakrzewski, Pawel
AU - Kim, Won-Keun
AU - Maisano, Francesco
AU - de Backer, Ole
AU - Dudek, Dariusz
PY - 2024/6/17
Y1 - 2024/6/17
N2 - BACKGROUND: Coronary access (CA) is a major concern in redo-transcatheter aortic valve implantation (TAVI) for failing supra-annular self-expanding transcatheter aortic valves (TAVs).AIMS: This ex vivo study evaluated the benefit of leaflet splitting (LS) on subsequent CA after redo-TAVI in anatomies deemed at high risk of unfeasible CA.METHODS: Ex vivo, patient-specific models were printed three-dimensionally. Index TAVI was performed using ACURATE neo2 or Evolut PRO (TAV-1) at the standard implant depth and with different degrees of commissural misalignment (CMA). Redo-TAVI was performed using the balloon-expandable SAPIEN 3 Ultra (TAV-2) at different implant depths with commissural alignment. Selective CA was attempted for each configuration before and after LS in a pulsatile flow simulator. The leaflet splay area was assessed on the bench.RESULTS: In matched comparisons of 128 coronary cannulations across 64 redo-TAVI configurations, the overall feasibility of CA significantly increased after LS (60.9% vs 18.7%; p<0.001). The effect of LS varied according to the sinotubular junction height, TAV-1 design, TAV-1 CMA, and TAV-2 implant depth, given TAV-2 alignment. LS enabled CA for up to CMA 45° with the ACURATE neo2 TAV-1 and up to CMA 30° with the Evolut PRO TAV-1. The combination of LS and a low TAV-2 implant provided the highest feasibility of CA after redo-TAVI. The leaflet splay area ranged from 25.60 mm2 to 37.86 mm2 depending on the TAV-1 platform and TAV-2 implant depth.CONCLUSIONS: In high-risk anatomies, LS significantly improves CA feasibility after redo-TAVI for degenerated supra-annular self-expanding platforms. Decisions on redo-TAVI feasibility should be carefully individualised, taking into account the expected benefit of LS on CA for each scenario.
AB - BACKGROUND: Coronary access (CA) is a major concern in redo-transcatheter aortic valve implantation (TAVI) for failing supra-annular self-expanding transcatheter aortic valves (TAVs).AIMS: This ex vivo study evaluated the benefit of leaflet splitting (LS) on subsequent CA after redo-TAVI in anatomies deemed at high risk of unfeasible CA.METHODS: Ex vivo, patient-specific models were printed three-dimensionally. Index TAVI was performed using ACURATE neo2 or Evolut PRO (TAV-1) at the standard implant depth and with different degrees of commissural misalignment (CMA). Redo-TAVI was performed using the balloon-expandable SAPIEN 3 Ultra (TAV-2) at different implant depths with commissural alignment. Selective CA was attempted for each configuration before and after LS in a pulsatile flow simulator. The leaflet splay area was assessed on the bench.RESULTS: In matched comparisons of 128 coronary cannulations across 64 redo-TAVI configurations, the overall feasibility of CA significantly increased after LS (60.9% vs 18.7%; p<0.001). The effect of LS varied according to the sinotubular junction height, TAV-1 design, TAV-1 CMA, and TAV-2 implant depth, given TAV-2 alignment. LS enabled CA for up to CMA 45° with the ACURATE neo2 TAV-1 and up to CMA 30° with the Evolut PRO TAV-1. The combination of LS and a low TAV-2 implant provided the highest feasibility of CA after redo-TAVI. The leaflet splay area ranged from 25.60 mm2 to 37.86 mm2 depending on the TAV-1 platform and TAV-2 implant depth.CONCLUSIONS: In high-risk anatomies, LS significantly improves CA feasibility after redo-TAVI for degenerated supra-annular self-expanding platforms. Decisions on redo-TAVI feasibility should be carefully individualised, taking into account the expected benefit of LS on CA for each scenario.
KW - Humans
KW - Transcatheter Aortic Valve Replacement/methods
KW - Aortic Valve/surgery
KW - Heart Valve Prosthesis
KW - Prosthesis Design
KW - Printing, Three-Dimensional
KW - Aortic Valve Stenosis/surgery
KW - Coronary Vessels/surgery
KW - Prosthesis Failure
KW - Reoperation
KW - coronary artery disease
KW - TAVI
KW - aortic stenosis
KW - valve-in-valve
UR - https://www.scopus.com/pages/publications/85196549303
U2 - 10.4244/EIJ-D-24-00107
DO - 10.4244/EIJ-D-24-00107
M3 - Journal article
C2 - 38887883
SN - 1774-024X
VL - 20
SP - e770-e780
JO - EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
JF - EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
IS - 12
ER -