Abstract
BACKGROUND: Antitachycardia pacing (ATP) is an effective treatment for ventricular tachycardia (VT). We evaluated the efficacy of different ATP programs based on a large remote monitoring data set from patients with implantable cardioverter-defibrillators (ICDs).
METHODS: A dataset from 18,679 ICD patients was used to evaluate the first delivered ATP treatment. We considered all device programs that were used for at least 50 patients, leaving us with 7 different programs and a total of 32,045 episodes. We used the two-proportions z-test (α = 0.01) to compare the probability of success and the probability for acceleration in each group with the corresponding values of the default setting.
RESULTS: Overall, the first ATP treatment terminated in 78.4%-97.5% of episodes with slow VT and 81.5%-91.1% of episodes with fast VT. The default setting of the ATP programs with the number of sequences S = 3 was applied to treat 30.1% of the slow and 36.6% of the fast episodes. Reducing the maximum number of sequences to S = 2 decreased the success rate for slow VT (P < 0.0001, h = 0.38), while the setting S = 4 resulted in the highest success rate of 97.5% (P < 0.0001, h = 0.27).
CONCLUSION: While the default programs performed well, we found that increasing the number of sequences from 3 to 4 was a promising option to improve the overall ATP performance.
| Originalsprog | Engelsk |
|---|---|
| Tidsskrift | PLoS One |
| Vol/bind | 14 |
| Udgave nummer | 8 |
| Sider (fra-til) | e0219533 |
| ISSN | 1932-6203 |
| DOI | |
| Status | Udgivet - 2019 |
| Udgivet eksternt | Ja |
Fingeraftryk
Dyk ned i forskningsemnerne om 'Impact of device programming on the success of the first anti-tachycardia pacing therapy: An anonymized large-scale study'. Sammen danner de et unikt fingeraftryk.Citationsformater
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS