TY - JOUR
T1 - Glucose and glycerol concentrations and their tracer enrichment measurements using liquid chromatography tandem mass spectrometry
AU - Bornø, Andreas
AU - Foged, Lene
AU - van Hall, Gerrit
N1 - Copyright © 2014 John Wiley & Sons, Ltd.
PY - 2014/10
Y1 - 2014/10
N2 - The present study describes a new liquid chromatography tandem mass spectrometry method for high-throughput quantification of glucose and glycerol in human plasma using stable isotopically labeled internal standards and is suitable for simultaneous measurements of glucose and glycerol enrichments in connection to in vivo metabolic studies investigating glucose turnover and lipolytic rate. Moreover, in order to keep up with this new fast analysis, simple derivatization procedures have been developed. Prior to analysis, glucose and glycerol were derivatized using benzoyl chloride in order to form benzoylated derivatives via new simplified fast procedures. For glucose, two internal standards were evaluated, [U-(13) C(6)]glucose and [U-(13) C(6), D(7)]glucose, and for glycerol, [U-(13) C(3), D(8)]glycerol was used. The method was validated by means of calibration curves, quality control samples, and plasma samples spiked with [6,6-D(2)]glucose, [U-(13) C(6)]glucose, and [1,1,2,3,3-D(5)]glycerol in order to test accuracy, precision, and recovery of the method. Moreover, post preparative and freeze-thaw sample stability were tested. The correlation of calibration curves for the glucose concentration were r(2) = 0.9998 for [U-(13) C(6)]glucose and r(2) = 0.9996 for [U-(13) C(6), D(7)]glucose, and r(2) = 0.9995 for the glycerol concentration. Interday accuracy for glucose using [U-(13) C(6)]glucose and glycerol determined in spiked plasma were respectively 103.5% and 106.0%, and the coefficients of variation were 2.0% and 9.7%, respectively. After derivatization, plasma samples were stable for at least 14 days. In conclusion, we have developed and validated a novel, accurate, and sensitive high-throughput liquid chromatography tandem mass spectrometry method for simultaneous determination of glucose and glycerol concentrations and enrichment of infused tracers most commonly used in human metabolic kinetic studies.
AB - The present study describes a new liquid chromatography tandem mass spectrometry method for high-throughput quantification of glucose and glycerol in human plasma using stable isotopically labeled internal standards and is suitable for simultaneous measurements of glucose and glycerol enrichments in connection to in vivo metabolic studies investigating glucose turnover and lipolytic rate. Moreover, in order to keep up with this new fast analysis, simple derivatization procedures have been developed. Prior to analysis, glucose and glycerol were derivatized using benzoyl chloride in order to form benzoylated derivatives via new simplified fast procedures. For glucose, two internal standards were evaluated, [U-(13) C(6)]glucose and [U-(13) C(6), D(7)]glucose, and for glycerol, [U-(13) C(3), D(8)]glycerol was used. The method was validated by means of calibration curves, quality control samples, and plasma samples spiked with [6,6-D(2)]glucose, [U-(13) C(6)]glucose, and [1,1,2,3,3-D(5)]glycerol in order to test accuracy, precision, and recovery of the method. Moreover, post preparative and freeze-thaw sample stability were tested. The correlation of calibration curves for the glucose concentration were r(2) = 0.9998 for [U-(13) C(6)]glucose and r(2) = 0.9996 for [U-(13) C(6), D(7)]glucose, and r(2) = 0.9995 for the glycerol concentration. Interday accuracy for glucose using [U-(13) C(6)]glucose and glycerol determined in spiked plasma were respectively 103.5% and 106.0%, and the coefficients of variation were 2.0% and 9.7%, respectively. After derivatization, plasma samples were stable for at least 14 days. In conclusion, we have developed and validated a novel, accurate, and sensitive high-throughput liquid chromatography tandem mass spectrometry method for simultaneous determination of glucose and glycerol concentrations and enrichment of infused tracers most commonly used in human metabolic kinetic studies.
KW - Blood Glucose/analysis
KW - Chromatography, Liquid/methods
KW - Glycerol/blood
KW - High-Throughput Screening Assays
KW - Humans
KW - Isotope Labeling/methods
KW - Linear Models
KW - Lipolysis
KW - Metabolomics
KW - Plasma/metabolism
KW - Reproducibility of Results
KW - Sensitivity and Specificity
KW - Tandem Mass Spectrometry/methods
U2 - 10.1002/jms.3407
DO - 10.1002/jms.3407
M3 - Journal article
C2 - 25303387
SN - 1076-5174
VL - 49
SP - 980
EP - 988
JO - Journal of mass spectrometry : JMS
JF - Journal of mass spectrometry : JMS
IS - 10
ER -