Effects of imaging gradients in sequences with varying longitudinal storage time-Case of diffusion exchange imaging

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Quantification of cerebral perfusion and cerebrovascular reserve using Turbo-QUASAR arterial spin labeling MRI

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Coil profile estimation strategies for parallel imaging with hyperpolarized 13 C MRI

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Gamma-aminobutyric acid edited echo-planar spectroscopic imaging (EPSI) with MEGA-sLASER at 7T

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Comparison of prospective head motion correction with NMR field probes and an optical tracking system

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  5. Improved calculation of the equilibrium magnetization of arterial blood in arterial spin labeling

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

PURPOSE: To illustrate the potential bias caused by imaging gradients in correlation MRI sequences using longitudinal magnetization storage (LS) and examine the case of filter exchange imaging (FEXI) yielding maps of the apparent exchange rate (AXR).

METHODS: The effects of imaging gradients in FEXI were observed on yeast cells. To analyze the AXR bias, signal evolution was calculated by applying matrix exponential operators.

RESULTS: A sharp threshold for the slice thickness was identified, below which the AXR is increasingly underestimated. The bias can be understood in terms of an extended low-pass diffusion filtering during the LS interval, which is more pronounced at lower exchange rates. For a total exchange rate constant larger than 1 s(-1) , the AXR bias is expected to be negligible when slices thicker than 2.5 mm are used.

CONCLUSION: In correlation experiments like FEXI, relying on LS with variable duration, imaging gradients may cause disrupting effects that cannot be easily mitigated and should be carefully considered for unbiased results. In typical clinical applications of FEXI, the imaging gradients are expected to cause a negligible AXR bias. However, the AXR bias may be significant in preclinical settings or whenever thin imaging slices are used. Magn Reson Med, 2017. © 2017 International Society for Magnetic Resonance in Medicine.