Early micromovement of the Articular Surface Replacement (ASR) femoral component: Two-year radiostereometry results

J O Penny, M Ding, Jens-Erik Varmarken, O Ovesen, S Overgaard

10 Citationer (Scopus)

Abstract

Radiostereometric analysis (RSA) can detect early micromovement in unstable implant designs which are likely subsequently to have a high failure rate. In 2010, the Articular Surface Replacement (ASR) was withdrawn because of a high failure rate. In 19 ASR femoral components, the mean micromovement over the first two years after implantation was 0.107 mm (sd 0.513) laterally, 0.055 mm (sd 0.204) distally and 0.150 mm (sd 0.413) anteriorly. The mean backward tilt around the x-axis was -0.08° (sd 1.088), mean internal rotation was 0.165° (sd 0.924) and mean varus tilt 0.238° (sd 0.420). The baseline to two-year varus tilt was statistically significant from zero movement, but there was no significant movement from one year onwards. We conclude that the ASR femoral component achieves initial stability and that early migration is not the mode of failure for this resurfacing arthroplasty.
OriginalsprogEngelsk
TidsskriftBone and Joint Journal
Vol/bind94
Udgave nummer10
Sider (fra-til)1344-50
Antal sider7
ISSN2049-4394
DOI
StatusUdgivet - 2012
Udgivet eksterntJa

Fingeraftryk

Dyk ned i forskningsemnerne om 'Early micromovement of the Articular Surface Replacement (ASR) femoral component: Two-year radiostereometry results'. Sammen danner de et unikt fingeraftryk.