Forskning
Udskriv Udskriv
Switch language
Region Hovedstaden - en del af Københavns Universitetshospital
Udgivet

Contact dermatitis caused by glucose sensors, insulin pumps, and tapes: Results from a 5-year period

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Degree of employment, sick leave, and costs following notification of occupational contact dermatitis-A register-based study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Patch test results with the European baseline series and additions thereof in the ESSCA network, 2015-2018

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Natural ingredients in cosmetic products - a suggestion for a screening series for skin allergy

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Degree of employment, sick leave, and costs following notification of occupational contact dermatitis-A register-based study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Clinical manifestations and impact on daily life of allergy to polyethylene glycol (PEG) in ten patients

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Patch test results with the European baseline series and additions thereof in the ESSCA network, 2015-2018

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer

BACKGROUND: The number of patients with contact dermatitis from glucose sensors and insulin infusion sets is increasing. Patch testing is challenging because of a lack of information concerning the constituent materials of medical devices.

OBJECTIVES: To report on products and causes of allergic reactions to glucose sensors or insulin infusion sets over a 5-year period and suggest a short screening series.

METHODS: Analysis of patch test data from consecutive patients suspected of allergic contact dermatitis (ACD) to glucose sensors and/or insulin infusion sets from 2015-2019.

RESULTS: Patient numbers increased from 4 to 15 per year; 30/38 (78.9%) were children. In 29 (76.3%), a diagnosis of allergic/probable ACD was established, mostly due to the tapes of the device or allergens in these tapes (n = 23) followed by allergens in the device housing (n = 10). Isobornyl acrylate, abitol, and colophonium were the most common allergens. Information from manufacturers was often difficult to obtain and, if accessible, inadequate. For this reason, the diagnosis was delayed for more than 1.5 years in 12 (31%) patients.

CONCLUSIONS: The increasing number of patients, mostly children, with ACD from devices used in treatment of type 1 diabetes demonstrates the importance of this problem. Allergies can easily be overlooked, due to the lack of mandatory labeling of the constituent materials of the devices.