TY - JOUR
T1 - Complications of implantable cardioverter-defibrillator treatment in arrhythmogenic right ventricular cardiomyopathy
AU - Christensen, Alex Hørby
AU - Platonov, Pyotr G
AU - Svensson, Anneli
AU - Jensen, Henrik K
AU - Rootwelt-Norberg, Christine
AU - Dahlberg, Pia
AU - Madsen, Trine
AU - Frederiksen, Tanja Charlotte
AU - Heliö, Tiina
AU - Haugaa, Kristina H
AU - Bundgaard, Henning
AU - Svendsen, Jesper H
N1 - Published on behalf of the European Society of Cardiology. All rights reserved. © The Author(s) 2021. For permissions, please email: [email protected].
PY - 2022/2/2
Y1 - 2022/2/2
N2 - AIMS: Treatment with implantable cardioverter-defibrillators (ICD) is a cornerstone for prevention of sudden cardiac death in arrhythmogenic right ventricular cardiomyopathy (ARVC). We aimed at describing the complications associated with ICD treatment in a multinational cohort with long-term follow-up.METHODS AND RESULTS: The Nordic ARVC registry was established in 2010 and encompasses a large multinational cohort of ARVC patients, including their clinical characteristics, treatment, and events during follow-up. We included 299 patients (66% males, median age 41 years). During a median follow-up of 10.6 years, 124 (41%) patients experienced appropriate ICD shock therapy, 28 (9%) experienced inappropriate shocks, 82 (27%) had a complication requiring surgery (mainly lead-related, n = 75), and 99 (33%) patients experienced the combined endpoint of either an inappropriate shock or a surgical complication. The crude rate of first inappropriate shock was 3.4% during the first year after implantation but decreased after the first year and plateaued over time. Contrary, the risk of a complication requiring surgery was 5.5% the first year and remained high throughout the study period. The combined risk of any complication was 7.9% the first year. In multivariate cox regression, presence of atrial fibrillation/flutter was a risk factor for inappropriate shock (P < 0.05), whereas sex, age at implant, and device type were not (all P > 0.05).CONCLUSION: Forty-one percent of ARVC patients treated with ICD experienced potentially life-saving ICD therapy during long-term follow-up. A third of the patients experienced a complication during follow-up with lead-related complications constituting the vast majority.
AB - AIMS: Treatment with implantable cardioverter-defibrillators (ICD) is a cornerstone for prevention of sudden cardiac death in arrhythmogenic right ventricular cardiomyopathy (ARVC). We aimed at describing the complications associated with ICD treatment in a multinational cohort with long-term follow-up.METHODS AND RESULTS: The Nordic ARVC registry was established in 2010 and encompasses a large multinational cohort of ARVC patients, including their clinical characteristics, treatment, and events during follow-up. We included 299 patients (66% males, median age 41 years). During a median follow-up of 10.6 years, 124 (41%) patients experienced appropriate ICD shock therapy, 28 (9%) experienced inappropriate shocks, 82 (27%) had a complication requiring surgery (mainly lead-related, n = 75), and 99 (33%) patients experienced the combined endpoint of either an inappropriate shock or a surgical complication. The crude rate of first inappropriate shock was 3.4% during the first year after implantation but decreased after the first year and plateaued over time. Contrary, the risk of a complication requiring surgery was 5.5% the first year and remained high throughout the study period. The combined risk of any complication was 7.9% the first year. In multivariate cox regression, presence of atrial fibrillation/flutter was a risk factor for inappropriate shock (P < 0.05), whereas sex, age at implant, and device type were not (all P > 0.05).CONCLUSION: Forty-one percent of ARVC patients treated with ICD experienced potentially life-saving ICD therapy during long-term follow-up. A third of the patients experienced a complication during follow-up with lead-related complications constituting the vast majority.
UR - http://www.scopus.com/inward/record.url?scp=85125376780&partnerID=8YFLogxK
U2 - 10.1093/europace/euab112
DO - 10.1093/europace/euab112
M3 - Journal article
C2 - 34279601
SN - 1099-5129
VL - 24
SP - 306
EP - 312
JO - Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
JF - Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
IS - 2
ER -