Comparison of prospective head motion correction with NMR field probes and an optical tracking system

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Quantification of cerebral perfusion and cerebrovascular reserve using Turbo-QUASAR arterial spin labeling MRI

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Coil profile estimation strategies for parallel imaging with hyperpolarized 13 C MRI

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Gamma-aminobutyric acid edited echo-planar spectroscopic imaging (EPSI) with MEGA-sLASER at 7T

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Improved calculation of the equilibrium magnetization of arterial blood in arterial spin labeling

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  5. Effects of imaging gradients in sequences with varying longitudinal storage time-Case of diffusion exchange imaging

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Value and limitations of intracranial recordings for validating electric field modeling for transcranial brain stimulation

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Accessibility of cortical regions to focal TES: Dependence on spatial position, safety, and practical constraints

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Electric field simulations for transcranial brain stimulation using FEM: an efficient implementation and error analysis

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Blinding is compromised for transcranial direct current stimulation at 1 mA for 20 minutes in young healthy adults

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  • Martin Eschelbach
  • Ali Aghaeifar
  • Jonas Bause
  • Jonas Handwerker
  • Jens Anders
  • Eva-Maria Engel
  • Axel Thielscher
  • Klaus Scheffler
Vis graf over relationer

PURPOSE: The aim of this study was to compare prospective head motion correction and motion tracking abilities of two tracking systems: Active NMR field probes and a Moiré phase tracking camera system using an optical marker.

METHODS: Both tracking systems were used simultaneously on human subjects. The prospective head motion correction was compared in an MP2RAGE and a gradient echo sequence. In addition, the motion tracking trajectories for three subjects were compared against each other and their correlation and deviations were analyzed.

RESULTS: With both tracking systems motion artifacts were visibly reduced. The precision of the field probe system was on the order of 50 µm for translations and 0.03° for rotations while the camera's was approximately 5 µm and 0.007°. The comparison of the measured trajectories showed close correlation and an average absolute deviation below 500 µm and 0.5°.

CONCLUSION: This study presents the first in vivo comparison between NMR field probes and Moiré phase tracking. For the gradient echo images, the field probes had a similar motion correction performance as the optical tracking system. For the MP2RAGE measurement, however, the camera yielded better results. Still, both tracking systems substantially decreased image artifacts in the presence of subject motion. Thus, the motion tracking modality should be chosen according to the specific requirements of the experiment while considering the desired image resolution, refresh rate, and head coil constraints.

OriginalsprogEngelsk
TidsskriftMagnetic Resonance in Medicine
Vol/bind81
Udgave nummer1
Sider (fra-til)719-729
Antal sider11
ISSN0740-3194
DOI
StatusUdgivet - jan. 2019