Forskning
Udskriv Udskriv
Switch language
Region Hovedstaden - en del af Københavns Universitetshospital
Udgivet

Comparison of INVOS 5100C and Nonin SenSmart X-100 oximeter performance in preterm infants with spontaneous apnea

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Gestational age-dependent development of the neonatal metabolome

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Changes in pulmonary oxygen content are detectable with laser absorption spectroscopy: proof of concept in newborn piglets

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Gut transit time, using radiological contrast imaging, to predict early signs of necrotizing enterocolitis

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Does height and IGF-I determine pubertal timing in girls?

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Central data monitoring in the multicentre randomised SafeBoosC-III trial - a pragmatic approach

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. The effect of early probiotic exposure on the preterm infant gut microbiome development

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Extremely Preterm Infant Admissions Within the SafeBoosC-III Consortium During the COVID-19 Lockdown

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Aerosol generation by respiratory support of neonates may be low

    Publikation: Bidrag til tidsskriftKommentar/debatForskningpeer review

  5. Antibiotic regimens for early-onset neonatal sepsis

    Publikation: Bidrag til tidsskriftReviewpeer review

Vis graf over relationer

BACKGROUND: Tissue oximeters are not interchangeable. Two instruments with sensors dedicated to preterm infants-INVOS 5100C and Nonin SenSmart X-100-have not yet been compared.

METHODS: By measuring cerebral oxygenation in ten preterm infants with spontaneous apneic episodes defined by pulse oximeter readings (SpO2) below 80%, as well as tissue oxygenation during vascular occlusion on the forearm of ten adults, simultaneously we compared performance in the hypoxic range.

RESULTS: We found the mean conversion equations to be StO2,SenSmart X-100 = 0.34 × StO2,INVOS 5100C + 44.8% during apnea in infants and StO2,SenSmart X-100 = 0.59 × StO2,INVOS 5100C + 34.4% during vascular occlusion. The individual regressions displayed large and statistically significant variations in both infants and adults. In three infants the INVOS sensor showed very little reaction to decreases in SpO2.

CONCLUSIONS: These findings confirm that different NIRS devices give very different estimates when the oxygenation is low. The large variation when compared to SpO2 suggest that the sensor placement is very important in preterm infants.

OriginalsprogEngelsk
TidsskriftPediatric Research
Vol/bind87
Udgave nummer7
Sider (fra-til)1244-1250
Antal sider7
ISSN0031-3998
DOI
StatusUdgivet - jun. 2020

ID: 62426992