Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites

Isabelle Isa Kristin Steineck, Zeinab Mahmoudi, Ajenthen Ranjan, Signe Schmidt, John Bagterp Jorgensen, Kirsten Nørgaard

30 Citationer (Scopus)

Abstract

Background: The aim was to compare the accuracy of the Dexcom® G4 Platinum continuous glucose monitor (CGM) sensor inserted on the upper arm and the abdomen in adults. Methods: Fourteen adults with type 1 diabetes wore two CGMs, one placed on the upper arm and one placed on the abdomen. Three in-clinic visits of 5 h with YSI (2300 STAT, Yellow Springs Instrument) measurements as comparator were performed. Each visit was followed by 4 days with seven-point self-monitoring of blood glucose (SMBG) in free-living conditions. Accuracy analyses on the paired CGM-YSI and CGM-SMBG measurements of the two CGM sensors were performed. Results: Using YSI as comparator, the overall Mean Absolute Relative Difference (MARD) for the CGMabd was 12.3% and CGMarm was 12.0%. The percentage of the CGM measurements in zone A of Clarke error grid analysis for the CGMabd was 85.6% and CGMarm was 86.0%. The hypoglycemia sensitivity for the CGMabd and CGMarm was 69.3%. Using SMBG as comparator, the overall MARD for the CGMabd was 12.5% and CGMarm was 12.0%. The percentage of the CGM measurements in zone A for the CGMabd was 84.1% and the CGMarm was 85.0%. The hypoglycemia sensitivity for the CGMabd was 60.0% and the CGMarm was 71.1%. All the P-values from the comparisons between the accuracy of CGMabd and CGMarm were >0.05. Conclusion: The accuracy of a Dexcom G4 Platinum CGM sensor placed on the upper arm was not different from the accuracy of the sensor placed on the abdomen in adults with type 1 diabetes.

OriginalsprogEngelsk
TidsskriftDiabetes Technology and Therapeutics
Vol/bind21
Udgave nummer5
Sider (fra-til)295-302
Antal sider8
ISSN1520-9156
DOI
StatusUdgivet - 1 maj 2019

Fingeraftryk

Dyk ned i forskningsemnerne om 'Comparison of Continuous Glucose Monitoring Accuracy Between Abdominal and Upper Arm Insertion Sites'. Sammen danner de et unikt fingeraftryk.

Citationsformater