Catheter-based flow measurements in hemodialysis fistulas - Bench testing and clinical performance

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

DOI

  1. Surveillance for hemodialysis access stenosis: usefulness of ultrasound vector volume flow

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Rating scale for the assessment of competence in ultrasound-guided peripheral vascular access - a Delphi Consensus Study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  3. Current use of ultrasound for central vascular access in children and infants in the Nordic countries--a cross-sectional study

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  4. Blood flow measurements during hemodialysis vascular access interventions - Catheter-based thermodilution or Doppler ultrasound?

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  5. Endovascular treatment of hemodialysis arteriovenous fistulas: is immediate post-interventional blood flow a predictor of patency

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  1. Reliable and valid assessment of procedural skills in resuscitative endovascular balloon occlusion of the aorta (REBOA)

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

  2. Dual-energy CT

    Publikation: Bidrag til tidsskriftReviewForskningpeer review

  3. Endovascular Treatment of Chronic and Acute on Chronic Mesenteric Ischaemia: Results From a National Cohort of 245 Cases

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Vis graf over relationer
Purpose: The purpose of this study was to perform bench and clinical testing of a catheter-based intravascular system capable of measuring blood flow in hemodialysis vascular accesses during endovascular procedures. Methods: We tested the Transonic ReoCath Flow Catheter System which uses the thermodilution method. A simulated vascular access model was constructed for the bench test. In total, 1960 measurements were conducted and the results were used to determine the accuracy and precision of the catheters, the effects of external factors (e.g., catheter placement, injection duration), and to test for systematic bias. In the clinical study, two interventional radiologists conducted a total of 250 measurements in 14 patients with arteriovenous fistulas to determine clinical precision and enable testing for bias between measurers.Results: Accuracy was excellent for both catheters with a high level of agreement between results from the ReoCath Flow Catheter System and the reference flowmeter. Clinical precision, expressed as the mean coefficient of variation, was 5.9% and 4.7% for the antegrade and retrograde catheters, respectively. Flow measurements were significantly affected by the distance between a stenosis and the tip of a retrograde catheter with the effect being proportional to the degree of stenosis. There was no systematic bias between measurers. Conclusions: The Reocath Flow Catheter System was found to be accurate and precise. Reliable results require careful attention to catheter placement. Blood flow measurements provide unique information on the hemodynamic status of a vascular access and have the potential to optimize results of interventions.
OriginalsprogEngelsk
TidsskriftThe journal of vascular access
Vol/bind13
Udgave nummer1
Sider (fra-til)45-50
Antal sider6
DOI
StatusUdgivet - 2012

ID: 33173593