Blood flow measurements during hemodialysis vascular access interventions - Catheter-based thermodilution or Doppler ultrasound?

Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

Harvard

APA

CBE

MLA

Vancouver

Author

Bibtex

@article{d9b840ae4cfa49d192a844326a48067b,
title = "Blood flow measurements during hemodialysis vascular access interventions - Catheter-based thermodilution or Doppler ultrasound?",
abstract = "Purpose: To test the clinical performance of catheter-based thermodilution and Doppler ultrasound of the feeding brachial artery for blood flow measurements during hemodialysis vascular access interventions.Methods: Thirty patients with arteriovenous fistulas who underwent 46 interventions had access blood flow measured before and after every procedure. Two methods, catheter-based thermodilution and Doppler ultrasound, were compared to the reference method of ultrasound dilution. Catheter-based thermodilution and Doppler ultrasound were performed during the endovascular procedures while flow by ultrasound dilution was determined within three days of the procedure. The methods were compared using regression analysis and tested for systematic bias. Results: Failure to position the thermodilutional catheter correctly was observed in 8 out of 46 (17%) pre-intervention measurements. Post-intervention measurements and ultrasound measurements were feasible in all patients. The average level of agreement was good when comparing catheter-based thermodilution to ultrasound dilution. However, blood flow by ultrasound dilution may differ by ±130 mL/min (±22%) at a flow level of 600 mL/min by thermodilution. Results from Doppler ultrasound displayed a moderate level of agreement on average when compared to ultrasound dilution. Blood flow by ultrasound dilution may differ by ±160 mL/min (±27%) at a flow level of 600 mL/min by Doppler ultrasound. No systematic bias was detected by either method.Conclusions: On average, results from catheter-based thermodilution were more in agreement with results from the ultrasound dilution technique compared to Doppler ultrasound. However, considering the cost and the high technical failure rate of the thermodilutional system, we recommend the use of ultrasound.",
author = "Heerwagen, {S{\o}ren T} and Hansen, {Marc A} and Schroeder, {Torben V} and Ladefoged, {S{\o}ren D} and Lars L{\"o}nn",
year = "2012",
doi = "10.5301/jva.5000007",
language = "English",
volume = "13",
pages = "145--151",
journal = "Journal of Vascular Access",
issn = "1129-7298",
publisher = "Wichtig Editore s.r.l",
number = "2",

}

RIS

TY - JOUR

T1 - Blood flow measurements during hemodialysis vascular access interventions - Catheter-based thermodilution or Doppler ultrasound?

AU - Heerwagen, Søren T

AU - Hansen, Marc A

AU - Schroeder, Torben V

AU - Ladefoged, Søren D

AU - Lönn, Lars

PY - 2012

Y1 - 2012

N2 - Purpose: To test the clinical performance of catheter-based thermodilution and Doppler ultrasound of the feeding brachial artery for blood flow measurements during hemodialysis vascular access interventions.Methods: Thirty patients with arteriovenous fistulas who underwent 46 interventions had access blood flow measured before and after every procedure. Two methods, catheter-based thermodilution and Doppler ultrasound, were compared to the reference method of ultrasound dilution. Catheter-based thermodilution and Doppler ultrasound were performed during the endovascular procedures while flow by ultrasound dilution was determined within three days of the procedure. The methods were compared using regression analysis and tested for systematic bias. Results: Failure to position the thermodilutional catheter correctly was observed in 8 out of 46 (17%) pre-intervention measurements. Post-intervention measurements and ultrasound measurements were feasible in all patients. The average level of agreement was good when comparing catheter-based thermodilution to ultrasound dilution. However, blood flow by ultrasound dilution may differ by ±130 mL/min (±22%) at a flow level of 600 mL/min by thermodilution. Results from Doppler ultrasound displayed a moderate level of agreement on average when compared to ultrasound dilution. Blood flow by ultrasound dilution may differ by ±160 mL/min (±27%) at a flow level of 600 mL/min by Doppler ultrasound. No systematic bias was detected by either method.Conclusions: On average, results from catheter-based thermodilution were more in agreement with results from the ultrasound dilution technique compared to Doppler ultrasound. However, considering the cost and the high technical failure rate of the thermodilutional system, we recommend the use of ultrasound.

AB - Purpose: To test the clinical performance of catheter-based thermodilution and Doppler ultrasound of the feeding brachial artery for blood flow measurements during hemodialysis vascular access interventions.Methods: Thirty patients with arteriovenous fistulas who underwent 46 interventions had access blood flow measured before and after every procedure. Two methods, catheter-based thermodilution and Doppler ultrasound, were compared to the reference method of ultrasound dilution. Catheter-based thermodilution and Doppler ultrasound were performed during the endovascular procedures while flow by ultrasound dilution was determined within three days of the procedure. The methods were compared using regression analysis and tested for systematic bias. Results: Failure to position the thermodilutional catheter correctly was observed in 8 out of 46 (17%) pre-intervention measurements. Post-intervention measurements and ultrasound measurements were feasible in all patients. The average level of agreement was good when comparing catheter-based thermodilution to ultrasound dilution. However, blood flow by ultrasound dilution may differ by ±130 mL/min (±22%) at a flow level of 600 mL/min by thermodilution. Results from Doppler ultrasound displayed a moderate level of agreement on average when compared to ultrasound dilution. Blood flow by ultrasound dilution may differ by ±160 mL/min (±27%) at a flow level of 600 mL/min by Doppler ultrasound. No systematic bias was detected by either method.Conclusions: On average, results from catheter-based thermodilution were more in agreement with results from the ultrasound dilution technique compared to Doppler ultrasound. However, considering the cost and the high technical failure rate of the thermodilutional system, we recommend the use of ultrasound.

U2 - 10.5301/jva.5000007

DO - 10.5301/jva.5000007

M3 - Journal article

C2 - 21786245

VL - 13

SP - 145

EP - 151

JO - Journal of Vascular Access

JF - Journal of Vascular Access

SN - 1129-7298

IS - 2

ER -

ID: 33173544