Abstract
PURPOSE: To compare automated refraction 1 week and 1 month after uncomplicated cataract surgery.
METHODS: In this prospective cohort study, we recruited patients in a 2-month period and included consecutive patients scheduled for bilateral small-incision phacoemulsification cataract surgery. The exclusion criteria were (i) corneal and/or retinal pathology that could lead to automated refraction miscalculation and (ii) surgery complications. Automated refraction was measured 1 week and 1 month after surgery.
RESULTS: Ninety-five patients met the in- and exclusion criteria and completed follow-up. The mean refractive shift in spherical equivalent was -0.02 dioptre (D) between 1 week and 1 month after surgery and not statistical significant (p = 0.78, paired t-test). The magnitude of refractive shift in either myopic or hyperopic direction was neither correlated to age, preoperative corneal astigmatism, axial length nor phacoemulsification energy used during surgery (p > 0.05 for all variables, regression analysis). The refractive target was missed with 1.0 D or more in 11 (12%) patients. In this subgroup, the mean refractive shift in spherical equivalent was 0.49 D between 1 week and 1 month after surgery with a trend towards statistical significance (p = 0.07, paired t-test). There was no difference in age, preoperative corneal astigmatism, axial length or phacoemulsification energy used during surgery compared to the remainder of the patients (p > 0.05 for all variables, unpaired t-test).
CONCLUSION: Automated refraction is stabile 1 week after uncomplicated cataract surgery, but there is a trend towards instability, if the refractive target is missed with 1.0 D or more.
Originalsprog | Engelsk |
---|---|
Tidsskrift | Acta Ophthalmologica |
Vol/bind | 96 |
Udgave nummer | 2 |
Sider (fra-til) | 149-153 |
Antal sider | 5 |
ISSN | 1755-375X |
DOI | |
Status | Udgivet - mar. 2018 |