Abstract
INTRODUCTION: After failed mask ventilation and tracheal intubation, guidelines issued by the Difficult Airway Society recommend placing a second generation supraglottic airway device to secure oxygenation. Ultimately, a secure airway can be obtained by tracheal intubation through the supraglottic airway device using a bronchoscope. In this randomised trial, we compared the AuraGain™ with the i-gel™ as conduit for bronchoscopic intubation under continuous oxygenation performed by a group of anaesthesiologists with variable experience in a general population of patients.
METHOD: We randomised one hundred patients who were equally allocated to flexible bronchoscopic intubation through the i-gel™ or the AuraGain™. In a random order, 25 anaesthesiologists each performed four intubations, two using the i-gel™ and two using the AuraGain™. Our primary outcome was 'total time for airway management'; i.e. total time from manually reaching the SAD to successful FBI confirmed at the end of the first inspiratory downstroke on the capnography curve.
RESULTS: In total, 87% (95% CI, 79%-92%) of the patients were successfully intubated through the allocated supraglottic airway device. There was no difference in total time for airway management between the i-gel™ and the AuraGain™ (199 vs. 227 s, p = .076). However, there was a difference in time for placement of the i-gel™, compared to the AuraGain™, (37 vs. 54 s, p < .001). There were nine failed intubations in the AuraGain™ group compared to four in the i-gel™ group (p = .147).
CONCLUSION: We found no difference in total time for airway management between using the i-gel™ and using the AuraGain™.
| Originalsprog | Engelsk |
|---|---|
| Tidsskrift | Acta Anaesthesiologica Scandinavica |
| Vol/bind | 66 |
| Udgave nummer | 5 |
| Sider (fra-til) | 589-597 |
| Antal sider | 9 |
| ISSN | 0001-5172 |
| DOI | |
| Status | Udgivet - maj 2022 |