T1 - An in-well direct lysis method for rapid detection of SARS-CoV-2 by real time RT-PCR in eSwab specimens
AU - Jørgensen, Rikke Lind
AU - Pedersen, Martin Schou
AU - Chauhan, Alisha Shazad
AU - Andreasson, Louise Munkholm
AU - Kristiansen, Gitte Qvist
AU - Lisby, Jan Gorm
AU - Rosenstierne, Maiken Worsøe
AU - Schønning, Kristian
N1 - Copyright © 2021 Elsevier B.V. All rights reserved.
PY - 2021/3
Y1 - 2021/3
N2 - BACKGROUND: Diagnostic real time reverse transcription PCR (rRT-PCR) is usually done using nucleic acid (NA) purified from the sample. In the SARS-CoV-2 pandemic reagents and utensils for NA purification has been in short supply. This has generated interest in methods that eliminate the need for NA purification.OBJECTIVES: To investigate if addition of detergent to rRT-PCR master mix (MM) enabled in-well direct lysis and detection of SARS-CoV-2 in clinical eSwab specimens.STUDY DESIGN: IGEPAL-CA-630 (IGEPAL) was added to SARS-CoV-2 MM to 0.3% final concentration and crude sample was added directly to the PCR well containing MM. Cycle of positivity (Cp) and categorical agreement was compared in samples tested in standard rRT-PCR after NA purification and in in-well lysis, direct rRT-PCR.RESULTS: In-well lysis direct rRT-PCR detected SARS-CoV-2 in 27/30 previously SARS-CoV-2+ samples with an average bias of 3.26 cycles (95%CI: 0.08-6.43 cycles). All 30 previously test negative samples remained negative when tested in in-well lysis, direct PCR.CONCLUSIONS: Supplementation of detergent to MM was shown to be useful for the detection of SARS CoV-2 in eSwab specimens (COPAN) by direct rRT-PCR without prior NA purification.
AB - BACKGROUND: Diagnostic real time reverse transcription PCR (rRT-PCR) is usually done using nucleic acid (NA) purified from the sample. In the SARS-CoV-2 pandemic reagents and utensils for NA purification has been in short supply. This has generated interest in methods that eliminate the need for NA purification.OBJECTIVES: To investigate if addition of detergent to rRT-PCR master mix (MM) enabled in-well direct lysis and detection of SARS-CoV-2 in clinical eSwab specimens.STUDY DESIGN: IGEPAL-CA-630 (IGEPAL) was added to SARS-CoV-2 MM to 0.3% final concentration and crude sample was added directly to the PCR well containing MM. Cycle of positivity (Cp) and categorical agreement was compared in samples tested in standard rRT-PCR after NA purification and in in-well lysis, direct rRT-PCR.RESULTS: In-well lysis direct rRT-PCR detected SARS-CoV-2 in 27/30 previously SARS-CoV-2+ samples with an average bias of 3.26 cycles (95%CI: 0.08-6.43 cycles). All 30 previously test negative samples remained negative when tested in in-well lysis, direct PCR.CONCLUSIONS: Supplementation of detergent to MM was shown to be useful for the detection of SARS CoV-2 in eSwab specimens (COPAN) by direct rRT-PCR without prior NA purification.
KW - COVID-19 Nucleic Acid Testing/methods
KW - COVID-19/diagnosis
KW - Detergents/chemistry
KW - Humans
KW - RNA, Viral/isolation & purification
KW - SARS-CoV-2/isolation & purification
KW - Specimen Handling/methods
KW - SARS-CoV-2
KW - IGEPAL-CA-630
KW - Direct PCR
KW - Crude lysate
UR - http://www.scopus.com/inward/record.url?scp=85099199858&partnerID=8YFLogxK
U2 - 10.1016/j.jviromet.2021.114062
DO - 10.1016/j.jviromet.2021.114062
M3 - Journal article
C2 - 33428990
SN - 0166-0934
VL - 289
SP - 1
EP - 3
JO - Journal of Virological Methods
JF - Journal of Virological Methods
M1 - 114062
ER -