A dosimetric comparison of real-time adaptive and non-adaptive radiotherapy: A multi-institutional study encompassing robotic, gimbaled, multileaf collimator and couch tracking

  • Emma Colvill
  • , Jeremy Booth
  • , Simeon Nill
  • , Martin Fast
  • , James Bedford
  • , Uwe Oelfke
  • , Mitsuhiro Nakamura
  • , Per Poulsen
  • , Esben Worm
  • , Rune Hansen
  • , Thomas Ravkilde
  • , Jonas Scherman Rydhög
  • , Tobias Pommer
  • , Per Munck Af Rosenschold
  • , Stephanie Lang
  • , Matthias Guckenberger
  • , Christian Groh
  • , Christian Herrmann
  • , Dirk Verellen
  • , Kenneth Poels
  • Lei Wang, Michael Hadsell, Thilo Sothmann, Oliver Blanck, Paul Keall
    92 Citationer (Scopus)

    Abstract

    PURPOSE: A study of real-time adaptive radiotherapy systems was performed to test the hypothesis that, across delivery systems and institutions, the dosimetric accuracy is improved with adaptive treatments over non-adaptive radiotherapy in the presence of patient-measured tumor motion.

    METHODS AND MATERIALS: Ten institutions with robotic(2), gimbaled(2), MLC(4) or couch tracking(2) used common materials including CT and structure sets, motion traces and planning protocols to create a lung and a prostate plan. For each motion trace, the plan was delivered twice to a moving dosimeter; with and without real-time adaptation. Each measurement was compared to a static measurement and the percentage of failed points for γ-tests recorded.

    RESULTS: For all lung traces all measurement sets show improved dose accuracy with a mean 2%/2mm γ-fail rate of 1.6% with adaptation and 15.2% without adaptation (p<0.001). For all prostate the mean 2%/2mm γ-fail rate was 1.4% with adaptation and 17.3% without adaptation (p<0.001). The difference between the four systems was small with an average 2%/2mm γ-fail rate of <3% for all systems with adaptation for lung and prostate.

    CONCLUSIONS: The investigated systems all accounted for realistic tumor motion accurately and performed to a similar high standard, with real-time adaptation significantly outperforming non-adaptive delivery methods.

    OriginalsprogEngelsk
    TidsskriftRadiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology
    Vol/bind119
    Udgave nummer1
    Sider (fra-til)159-65
    Antal sider7
    ISSN0167-8140
    DOI
    StatusUdgivet - apr. 2016

    Fingeraftryk

    Dyk ned i forskningsemnerne om 'A dosimetric comparison of real-time adaptive and non-adaptive radiotherapy: A multi-institutional study encompassing robotic, gimbaled, multileaf collimator and couch tracking'. Sammen danner de et unikt fingeraftryk.

    Citationsformater