TY - JOUR
T1 - 2-Year Clinical Outcomes of an Abluminal Groove-Filled Biodegradable-Polymer Sirolimus-Eluting Stent Compared With a Durable-Polymer Everolimus-Eluting Stent
AU - Xu, Bo
AU - Saito, Yuichi
AU - Baumbach, Andreas
AU - Kelbæk, Henning
AU - van Royen, Niels
AU - Zheng, Ming
AU - Morel, Marie-Angèle
AU - Knaapen, Paul
AU - Slagboom, Ton
AU - Johnson, Thomas W
AU - Vlachojannis, Georgios
AU - Arkenbout, Karin E
AU - Holmvang, Lene
AU - Janssens, Luc
AU - Brugaletta, Salvatore
AU - Naber, Christoph K
AU - Anderson, Richard
AU - Rittger, Harald
AU - Berti, Sergio
AU - Barbato, Emanuele
AU - Toth, Gabor G
AU - Maillard, Luc
AU - Valina, Christian
AU - Buszman, Paweł
AU - Thiele, Holge
AU - Schächinger, Volker
AU - Lansky, Alexandra
AU - Wijns, William
AU - TARGET AC Investigators
N1 - Copyright © 2019 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
PY - 2019/9/9
Y1 - 2019/9/9
N2 - OBJECTIVES: The aim of this study was to assess the 2-year clinical outcomes of the Firehawk stent (Shanghai MicroPort Medical Group, Shanghai, China), a novel abluminal groove-filled biodegradable-polymer sirolimus-eluting coronary stent, compared with XIENCE (Abbott Vascular, Santa Clara, California), a durable-polymer everolimus-eluting coronary stent.BACKGROUND: The long-term outcomes of the Firehawk stent have not been evaluated beyond 1 year in a randomized all-comers clinical trial.METHODS: The TARGET All Comers study is a prospective, multicenter, all-comers, randomized, noninferiority trial conducted in Europe. A total of 1,653 patients were randomly assigned to undergo implantation of either the Firehawk or the XIENCE stent. The primary endpoint was target lesion failure, a composite of cardiac death, target vessel myocardial infarction, or ischemia-driven target lesion revascularization.RESULTS: At 2-year follow-up, the incidence of target lesion failure was 8.7% in the Firehawk group versus 8.6% in the XIENCE group (p = 0.92). The event rates of individual components of the primary endpoint were comparable for the 2 groups. Landmark analyses between 1- and 2-year follow-up revealed no statistically significant difference of TLF for the Firehawk versus the XIENCE stent. Beyond 1 year, very late definite or probable stent thrombosis occurred in 3 patients (0.4%) in the Firehawk group and in 7 patients (0.9%) in the XIENCE group (p = 0.34).CONCLUSIONS: The 2-year follow-up of the TARGET All Comers study confirms comparable safety and efficacy profiles of the Firehawk and XIENCE stents.
AB - OBJECTIVES: The aim of this study was to assess the 2-year clinical outcomes of the Firehawk stent (Shanghai MicroPort Medical Group, Shanghai, China), a novel abluminal groove-filled biodegradable-polymer sirolimus-eluting coronary stent, compared with XIENCE (Abbott Vascular, Santa Clara, California), a durable-polymer everolimus-eluting coronary stent.BACKGROUND: The long-term outcomes of the Firehawk stent have not been evaluated beyond 1 year in a randomized all-comers clinical trial.METHODS: The TARGET All Comers study is a prospective, multicenter, all-comers, randomized, noninferiority trial conducted in Europe. A total of 1,653 patients were randomly assigned to undergo implantation of either the Firehawk or the XIENCE stent. The primary endpoint was target lesion failure, a composite of cardiac death, target vessel myocardial infarction, or ischemia-driven target lesion revascularization.RESULTS: At 2-year follow-up, the incidence of target lesion failure was 8.7% in the Firehawk group versus 8.6% in the XIENCE group (p = 0.92). The event rates of individual components of the primary endpoint were comparable for the 2 groups. Landmark analyses between 1- and 2-year follow-up revealed no statistically significant difference of TLF for the Firehawk versus the XIENCE stent. Beyond 1 year, very late definite or probable stent thrombosis occurred in 3 patients (0.4%) in the Firehawk group and in 7 patients (0.9%) in the XIENCE group (p = 0.34).CONCLUSIONS: The 2-year follow-up of the TARGET All Comers study confirms comparable safety and efficacy profiles of the Firehawk and XIENCE stents.
U2 - 10.1016/j.jcin.2019.05.001
DO - 10.1016/j.jcin.2019.05.001
M3 - Journal article
C2 - 31129092
SN - 1936-8798
VL - 12
SP - 1679
EP - 1687
JO - JACC: Cardiovascular Interventions
JF - JACC: Cardiovascular Interventions
IS - 17
ER -