Agreement Between Transcutaneous Monitoring and Arterial Blood Gases During COPD Exacerbation

Research output: Contribution to journalJournal articleResearchpeer-review

DOI

  1. Comorbidity Clusters and Healthcare Use in Individuals With COPD

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Socioeconomic Variations in Use of Prescription Medicines for COPD: A Register-Based Study

    Research output: Contribution to journalJournal articleResearchpeer-review

  1. Direct effects of mast cell proteases, tryptase and chymase, on bronchial epithelial integrity proteins and anti-viral responses

    Research output: Contribution to journalJournal articleResearchpeer-review

  2. Mast cell tryptase enhances wound healing by promoting migration in human bronchial epithelial cells

    Research output: Contribution to journalJournal articleResearchpeer-review

  3. Airway hyperresponsiveness to inhaled mannitol identifies a cluster of non-eosinophilic asthma patients with high symptom burden

    Research output: Contribution to journalJournal articleResearchpeer-review

  4. Waking Up in Pain: a prospective unselected cohort study of pain in 3702 patients immediately after surgery in the Danish Realm

    Research output: Contribution to journalJournal articleResearchpeer-review

View graph of relations

BACKGROUND: Transcutaneous measurements of CO2 and O2 ([Formula: see text], [Formula: see text]) are noninvasive and allow for continuous monitoring in adults with exacerbation of COPD, but substantial accuracy issues may exist. We investigated agreement between results of arterial blood gas analysis and transcutaneous measurements of CO2 and O2 in patients with COPD.

METHODS: Adult subjects were monitored after acute admission to a respiratory intermediate care unit or ICU due to exacerbation of COPD and with ongoing noninvasive ventilation or immediately following extubation. Monitored variables were continuous transcutaneous measurement and simultaneous routine arterial blood gas analysis. Agreement between measurements was assessed by calculating bias with 95% limits of agreement for single-point estimates of [Formula: see text] versus [Formula: see text] and versus [Formula: see text], and for changes in transcutaneous measurements between 2 time points ([Formula: see text] and [Formula: see text]). We considered limits of agreement within ± 7.5 mm Hg to be acceptable.

RESULTS: A total of 57 transcutaneous measurements were made in 20 subjects for comparison with concurrent arterial blood gas analysis at 36 time points. The bias (limits of agreement) for [Formula: see text] and [Formula: see text] was 2.5 mm Hg (-10.6 to 15.6 mm Hg) and 11.2 mm Hg (-28.2 to 50.6 mm Hg), respectively. The bias for [Formula: see text] and [Formula: see text] was 2.3 mm Hg (-3.8 to 8.3 mm Hg) and -5.3 mm Hg (-37.5 to 27 mm Hg), respectively.

CONCLUSIONS: [Formula: see text] and [Formula: see text] did not accurately reflect results from arterial blood gas analyses in this study of mostly hypercapnic subjects. Agreement between changes in CO2 during the monitoring period was acceptable, however, and transcutaneous monitoring may be used for continuous monitoring of [Formula: see text] in conjunction with arterial blood gas analysis for reference.

Original languageEnglish
JournalRespiratory Care
Volume66
Issue number10
Pages (from-to)1560-1566
Number of pages7
ISSN0020-1324
DOIs
Publication statusPublished - Oct 2021

    Research areas

  • Adult, Blood Gas Monitoring, Transcutaneous, Carbon Dioxide, Humans, Hypercapnia, Noninvasive Ventilation, Pulmonary Disease, Chronic Obstructive, Intensive care, Transcutaneous blood gas monitoring, Noninvasive ventilation, Respiratory insuffi-ciency, COPD

ID: 65947890